This unit consist of Optics, Temperature Sensor Signal amplifier, Processing circuit and LCD Display. The Optics collected the infrared energy emitted by object and focus onto the Sensor. Then the sensor translates the energy into an electricity signal. This signal will be turned out to be digital shown on the LCD after the signal amplifier and processing circuit.

Infrared thermometer possesses the following features:

1.Positioning laser
2.℃/℉ temperature unit conversion
3.K-type thermocouple
4.Data retention
5.Low battery indicator
6.Environmental humidity measurement
7.Dew point temperature measurement
8.Surface temperature measurement

LCD
Colorful LCD
Temperature
measurement range
-50℃~650℃ (-58℉~1202℉)
Temperature display
accuracy
0.1℃
Temperature
measurement error
0℃~650℃(32℉~1202℉):±1.5% rdg
-50℃~0℃(-58℉~32℉):±3℃(±5℉)
or±1.5℃(±2.7℉)Take the bigger value
K-type thermocouple
measurement range
-40℃~500℃ (-40℉ ~ 932℉)
Repeatability
1% rdg or 1℃ Take the bigger value
Reponse time
0.5s, 95% Response
Emissivity
0.10~1.00 Adjustable(preset as 0.95)
D:S
12:1
Response wavelength
5um~14 um
Laser
Ring light spot
Flashlight
UV light
Ambient temperature
measurement range
-20 ℃~60℃ (-4℉ ~ 140℉)
Ambient temperature

measurement accuracy

±1℃(2℉)
Relative humidity
measurement accuracy
±5% RH
Dew point temperature
measurement
-10℃~ 50℃(14℉~ 122℉):±1.5℃/3℉
Mildew alarm indicator
Batteries
1.5V AAA battery*2(No.7 battery)
Low power indicator
Low power indicator for power below 2.4V
Overload indicator
“Hi”/”Lo” displayed on LCD
Weight/dimensions
202g(including batteries) /108*49*177 mm

 

Specifications of K-type thermocouple
Measurement range
0℃~250℃ (300℃ for short time)
Measurement error
±2.5℃ or 0.75% rdg Take the bigger value
Thermal response time
<10s