This unit consist of Optics, Temperature Sensor Signal amplifier, Processing circuit and LCD Display.The Optics collected the infrared energy emitted by object and focus onto the Sensor. Then the sensor translates the energy into an electricity signal. This signal will be turned out to be digital shown on the LCD after the signal amplifier and processing circuit.
  • Infrared thermometer possesses the following features:

    1.Positioning laser
    2.Backlit screen
    3.Data retention
    4.High and low temperature alarm
    5.Max/Min display
    6.Low battery reminder
    7.Self-calibration
    8.℃/℉ conversion

LCD
Colorful LCD
Temperature
range
-50℃~600℃(-58℉~1112℉)
Temperature
measurement error
-50~0℃ ( -58 ~32℉):±3℃ (±5℉ )
0 ~600℃ (32 ~1112℉ ):±1.5% rdg
or±1.5℃(±2.7℉ )Take the bigger value
Rrepeatability
1% rdg or 1℃Take the bigger value
Reponse time
0.5s, 95% Response
Emissivity
0.10~1.00 Adjustable(preset as 0.95)
D:S
12:1
Response
wavelength
5μm~14μm
No operation
shutdown
About 7 seconds
Batteries
1.5V AAAbattery*2(No.7 battery)
Low power
indicator
Low power indicator for power below
2.5V
Overload
indicator
“Hi”/“Lo” displayed on LCD
Exceed upper/lower
limits of work
environment
“AH”/“AL”displayed on LCD
Working environment
temperature
0℃~40℃(32℉~104℉)
Storage
temperature
-10℃~60℃(14℉~140℉)